OmniOx units built with blower, heated humidifier, and built-in oxygen blender was developed as the world's first equipment hat integrates heated humidified high oxygen nasal flow cannula treatment with non-invasive positive- pressure ventilation mode as CPAP and bi-level, enabling safe ventilation treatment in various environments from hospital to home.
The Omniox HFT750 is an all-in-one non-invasive respiratory solution (HF, CPAP, bi-level), caring for all respiratory patients, except invasive ventilation cases.
The Omniox 750A is available with only High Flow Therapy software enabled. It can be upgraded to full spec Omniox 750A.
Features of Omniox 750
Display: 5" touch screen
Intuitive and user-friendly UI
Rapid response to patients’ conditions by monitoring their respiration rate, SpO2, and S/F Ratio
Supplying stable oxygen through the oxygen blending technology
Equipped with TSF (Tissue Saturation Fraction) function to automatically adjust FiO2 concentration by measuring the patient's SpO2.
Easily building the breathing circuit of a patient and applications
Specifications: 1) High Flow (HF) mode
High flow setup : 1 - 60 LPM
FiO2 setup : 21 - 100 %
Monitoring factors : PTIF, ROX Index, SpO2, S/F ratio, RR, PR
2) CPAP mode
CPAP setup : 4 - 20 cmH2O
PA (+) : Off, 1 - 3 cmH2O
PA (-) : OFF, -3 - -1 cmH2O
3) BiPAP mode
IPAP setup : 4 - 40 cmH2O ± (1.7 + 4 % of the set value) cmH2O
EPAP setup : 4 - 20 cmH2O ± (1.7 + 4 % of the set value) cmH2O
Respiratory Rate : Off, 2 - 60 BPM
Ins. time : 0.5 - 3.0sec
Remote monitoring of patients by linking with the central monitoring system